Model 6200

The Model 6200 is designed to test most helicopters, general aviation aircraft, and eVTOLs. It is used to perform leak checks and non-RVSM accuracy checks on the Pitot-Static systems – including altimeters, airspeed indicators and VSI’s – and can be done at low airspeeds. The Model 6200 includes a Remote unit with large keypad and display, a removable 8-hour battery (optional), wide operating temperature (optional), and “profiles” to accelerate testing. The unit can be controlled by an iPad or other tablets.

Request a quote?

STATIC PORT

Pressure Function

Range: 1.0 to 32 inHg

Resolution: 0.001 inHg

Accuracy: ±0.005 inHg

Units: inHg mbar, psi, mmHg, kpa

Pressure Rate

Range: 0 to 30 inHg/min

Resolution: 0.001 inHg/min

Accuracy: ±1% of rate

Units: pressure-units/min

Altitude Function

Range: -2000 ft to 50,000 ft

Resolution: 1 ft

Accuracy:

  • ± 5 ft @ 0 ft
  • ± 6.5 ft @ 10,000 ft
  • ± 10 ft @ 20,000 ft
  • ± 15 ft @ 35,000 ft
  • ± 30 ft @ 50,000 ft

Units: feet, meters

Climb Function

Range: 0 ft/min to ±10,000 ft/min

Resolution: 1 ft/min

Accuracy: ±1% of rate of climb

Units: ft/min, meters/sec

SYSTEM

Power Requirement

90-260 VAC, 47-440 Hz, 50 VA

External Interfaces

Wi-Fi, RS232 Serial, Encoder (optional)

Dimensions & weights

Main unit: 22” x 14” x 9” / 30 lbs

Remote unit: 6” x 6” x 1” / 1.8 lbs

Environmental Specs

Operating temp: 0°C to 50°C, –40°C to 50°C (optional)

Storage temp: –40°C to 75°C

Humidity: 5 to 100%

PITOT PORT

Pressure Function

Range: 0 to 15 inHg (differential)

Resolution: 0.001 inHg

Accuracy: ±0.003 inHg

Pt Units: inHg, mbar, psi, mmHg, kpa, EPR

Qc Units: inHg, mbar, psi, mmHg, kpa

Pressure Rate

Range: 0 to 30 inHg/min

Resolution: 0.001 inHg/min

Accuracy: ±1% of rate

Units: pressure-units/min (EPR in inHg/min)

Airspeed Function

Range: 0 to 500 knots

Resolution: 0.1 knots

Accuracy:

  • ± 1.0 knots @ 20 knots
  • ± 0.3 knots @ 50 knots
  • ± 0.2 knots @ 100 knots
  • ± 0.1 knots @ 200 knots
  • ± 0.04 knots @ 500 knots

Units: knots, Mach, km/hr, mph

Rate of change: 0 to 500 knots/min

Airspeed Rate

Range: 0 to 500 knots/min

Resolution: 0.1 knots/min (0.01 km/hr/sec)

Accuracy: ±2% of rate

Units: knots/min, km/hr/sec, mph/min

Rate of change: 0 to 500 knots/min

Mach Function

Range: 0.0 to 1.0 Mach

Resolution: 0.001 Mach

Accuracy: ±0.001 above 0.2 Mach

Laversab Inc.

Innovation is built into our culture
Founded in 1982, Laversab began as a key supplier to governments worldwide for high accuracy, pressure calibration instrumentation. As an audited and approved subcontractor to the United States Department of Defense, Laversab supplies highly rugged, military-specification products for the US Air Force and US Navy, as well as other government agencies. Today, most major aircraft manufacturers, airlines and certificated repair stations across the country and the world are also Laversab customers.